| Model Number |
L301 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003, which states that the voltage is too low for projected remaining capacity.It was reported that a request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended.This device remains in service.No adverse patient effects were reported.Additional information was received indicating that this device was explanted and replaced.No additional adverse patient effects were reported, and this device has not been returned.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|