It was reported that the patient was undergoing neurological electrical stimulation therapy for cancer tumors in the lungs via a vest-like system delivering electrical stimulation over a twelve hour period.This system was interfering with the implanted implantable cardioverter defibrillator (icd), resulting in a spike and out-of-range low right atrial (ra) lead impedance with noise and oversensing observed.The right ventricular (rv) lead exhibited high, varying, and low out-of-range defibrillation impedance, as well a spike, out-of-range high, and out-of-range low rv pacing impedance values.It was reported that all lead alerts and trends were directly associated with the patient¿s use of the neurological stimulation system, and electromagnetic interference (emi) was identified as the cause of the out-of-range measurements and oversensing.Once the patient¿s physician elected to suspend the cancer treatment, impedance measurements appeared to remain stable with no additional oversensing observed.The device currently appears to be functioning as programmed.The icd remains in use.No patient complications have been reported as a result of this event.
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|