Tablo instructions for use (ifu) warning and caution: check all bloodlines for leaks after the treatment has started.Keep access sites uncovered and monitored.Improper bloodline connections or needle dislodgements can result in excessive blood loss, serious injury, and death.Machine alarms may not occur in every blood loss situation.The actual device was not returned; however, a photograph was provided for evaluation.Visual inspection of the photo did confirm a cartridge leak.Investigation identified a mechanical problem that was traced to the manufacturing process as the most probable cause of the event.Appropriate manufacturing actions have been initiated.A review of the device history record (dhr) and production records for this lot did not identify any documented manufacturing nonconformances.
|