| Model Number |
470500-13 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the universal seal had an obstruction at the insertion point.The procedure was completing as planned with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The universal seal has been returned and evaluated by the failure analysis team.Failure analysis investigations replicated/confirmed the customer reported complaint.The seal was found to have torn at the septum, measuring approximate 3.50 mm in length.The component is damaged and there be missing material, was not returned with the accessory.The complaint was confirmed by failure analysis.The probable root cause of the torn seal is attributed to damage during various handling points, including installation or use.
|
| |
|