| Model Number |
L310 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this device was operating in safety mode.Technical services (ts) was consulted and found an unknown code and submitted a request for an engineering analysis.Ts recommended device replacement.The device was subsequently explanted, and a new device was successfully implanted.No additional adverse patient effects were reported.This device is no longer in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|