Model Number DIB00
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during the initial stages of implanting a single-piece preloaded monofocal intraocular lens (iol), a clear plastic particle was observed in the anterior chamber.The lens was not implanted, and the particle was retrieved.The reporter indicated that the particle was not believed to have originated from the iol.No patient complications or adverse effects were reported.A backup lens was used to complete the procedure.No further information was provided.
 
Additional Manufacturer Narrative
Section a2: date of birth, age: unknown/not provided.Section a3a, a3b: sex, gender: unknown/not provided.Section a4: weight: unknown/not provided.Section a5: ethnicity: unknown/not provided.Section a6: race: unknown/not provided.Section d6a: if implanted, give date: unknown/not provided; the extent of patient contact is unknown, however there is no indication that the lens was fully implanted.Section d6b: if explanted, give date: unknown/not provided; the extent of patient contact is unknown, however there is no indication that the lens was fully implanted, therefore, not explanted.Section b3: date of event: unknown; requested but not provided.The best estimate date is prior to jul 21, 2026 "[alert date]".Device evaluation: product evaluation was not performed because the product has not been received.The complaint issue reported could not be confirmed.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search in the complaint system revealed that no other complaints were received for this production order number.Conclusion: based on the investigation, no malfunction or product deficiency was identified.Attempts were made to obtain the missing information; however, no additional information has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
Intraocular Lens
Manufacturer (Section D)
AMO Puerto Rico Mfg. Inc.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
+1(714)247-8552
MDR Report Key50267130
Report Number3012236936-2026-000240
Device Sequence Number13406291
Product Code HQL
UDI-Device Identifier05050474731622
UDI-Public(01)05050474731622(17)281110
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDIB00
Device Catalogue NumberDIB00U0135
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2026
Initial Report FDA Received Date08/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-