| Model Number |
500004500500 |
| Medical Device Problem Code |
Labelling, Instructions for Use or Training Problem (1318)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medbank mini, user stated that they were unable to issue medication for the patient.The customer stated that this malfunction occurred while dispensing the medication.There were no adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 10-mar-2022 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the user was unable to issue medication.A technical support specialist (tss) assigned the appropriate drawer and reviewed the configuration.The tss then had the customer recreate the transaction, after which the medications were successfully issued.The issue was resolved.The system functioned as intended after the technical support specialist troubleshot the device.
|
| |
|
Search Alerts/Recalls
|