| Model Number |
IntelliVue Patient Monitor MX500 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device was providing inaccurate ecg data.The device was in use on a patient at the time of the event.There was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An analysis was performed by a philips field service engineer (fse).The fse determined that the device was providing poor ecg signal with artifacts even after replacing the electrodes.The fse determined that the ecg lead set needed to be replaced.Based on the results of the analysis, it was determined that the product has malfunctioned and the most likely cause of the reported problem was a faulty ecg lead set.The issue was resolved by replacing the ecg lead set cables