Model Number MODEL 250 |
Device Problems
Image Display Error/Artifact (1304); Failure to Power Up (1476)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/24/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that the physician's tablet would not power on.Once the tablet finally did power on, the screen was noted to be full of lines.The complaint tablet was received on (b)(6) 2015.The physician was provided a new programming tablet.Device analysis is underway but has not been completed to date.
|
|
Event Description
|
Product analysis was performed which showed the cause of the reported power on issues was associated with an upside down power button.Once the power button was installed correctly, no further anomalies associated with the tablet were identified.The dhr was reviewed and it was noted the device passed all functionality tests and inspection.
|
|
Manufacturer Narrative
|
The udi # was inadvertently left of the original mfr.Report; (b)(4).
|
|