| Model Number |
CH-S700-XZ-EA |
| Medical Device Problem Codes |
Degraded (1153); Communication or Transmission Problem (2896); Gas/Air Leak (2946); Optical Distortion (3000); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the camera head had distorted image.During the initial evaluation of the device, it was observed that smashed connector, failed leak test, image quality error (error e238) shown on monitor.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was not confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: smashed connector, failed leak test, image quality error (error e238) shown on monitor
a definitive root cause could not be identified.Based on the results of the investigation, the most probable cause of the reported malfunction could not be determined.Additionally, the cause of the malfunction was traced to the component failure
should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|