A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 21-mar-2019 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the system wrong dosage medications stocked on patient profile.A technical support specialist (tss) communicated with the pharmacy, which confirmed that the medication was stocked in the cabinet.The tss reviewed the patient's profile in myqlink (mql) and found that the patient had an order for the 10 mg tablet, while the cabinet was stocked with the 5 mg tablet.As a result, the medication did not appear on the patient's profile for issuing.The tss informed the customer that pharmacy would need to ensure the correct medication was available on the patient's profile and need to send the medications accordingly.The system functioned as intended after the technical support specialist assessed the device.
|