| Model Number |
LifeVest WCD 5100 System |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported a battery pack has bent pins identified after patient fitting.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of battery has been completed.The reported problem (battery pack has bent pins identified after to patient fitting) was confirmed.As received, the battery would not firmly latch into the monitor due to bent pins in the battery connector.There was no adverse event that occurred related to this issue.The root cause of the physical damage to the bent pins was unable to be positively identified.There were no adverse events associated with the damaged battery.
|
| |
|