It was reported that device embolization occurred.A left atrial appendage (laa) closure procedure was being performed, and a 35mm watchman flx pro closure device was selected to be used.The patient shape of the appendage was chicken wing.During the procedure, after attempting several closure devices, the third device met position, anchor, size and seal (pass) criteria.The specific compression was 20-26 percent; the left atrium mean pressure was two (2) and the patient received an iv fluid bolus.The physician released the closure device, and it immediately embolized.The physician unsuccessfully attempted to snare the closure device from the left atrium.The closure device traveled to the left ventricle and then into the ascending aorta.Vascular surgery performed a cutdown on the left femoral artery and the closure device was successful snared and removed from the ascending aorta.The physician suggested the volume status was the reason for smaller pre-ostium measurements.The physician used three (3) closure devices during this implant and was confident by the time the third closure device was used, the patient's volume should have stabilized.There is no plan to retry implant of the closure device.The patient was discharged.
|