Visual inspection
during the investigation, no significant deformations or damage of the valves was determined.Permeability test
a permeability test has indicated that the m.Blue has a blockage.Computer controlled test
to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in vertical/ or both the horizontal as well as the vertical positions.Because the valve is not permeable, a computer controlled test is not applicable.Adjustment test
the m.Blue valve was tested and is not adjustable throughout the normal range.Braking force and brake function test
the brake functionality test has shown that the brake function is operational.However, the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection
after dismantling of the valve, organic deposits were found.Results
according to the investigation results, a non-adjustability and a blockage of the m.Blue were detected.The visible deposits led to the visible deposits led to the aforementioned functional impairments.Deposits caused by substances naturally present in the patient's body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve.
|