The account reported finding the sample ¿no longer suctions.¿ medline quality received one unit; visual inspection found it appeared to be in used condition but no physical damage was present, and testing showed the sample turned on, the pressure observed at maximum function was greater than 600 mm hg, and the unit did create suction while running it.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|