• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Ireland Diagnostics Division Alinity I Free T4 Reagent Kit; Radioimmunoassay, free thyroxine

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Ireland Diagnostics Division Alinity I Free T4 Reagent Kit; Radioimmunoassay, free thyroxine Back to Search Results
Catalog Number 07P70-30
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed falsely depressed alinity i free t4 results generated for patient samples.The following data was provided (customer¿s reference range 0.75-1.22 ng/dl): sample id (b)(6), (41-year-old female) initial result = 0.69 ng/dl, repeat = 0.84 ng/dl.Sample id (b)(6), (39-year-old female) initial result = 0.68 ng/dl, repeat = 0.76 ng/dl.Sample id (b)(6), (36-year-old female) initial result = 0.70 ng/dl, repeat = 0.79 ng/dl.Sample id (b)(6), (b)(6) initial result = 0.74 ng/dl, repeat = 0.88 ng/dl.Sample id (b)(6), (62-year-old female) initial result = 0.73 ng/dl, repeat = 0.87 ng/dl.Sample id (b)(6), (50-year-old female) initial result = 0.67 ng/dl, repeat = 0.76 ng/dl.Sample id (b)(6), (45-year-old female) initial result = 0.73 ng/dl, repeat = 0.84 ng/dl.Sample id (b)(6), (20-year-old female) initial result = 0.66 ng/dl, repeat = 0.75 ng/dl.No impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A final report will be submitted when the evaluation is complete.Section a patient information: refer to section b5 for complete patient information.This report is being filed on an international product, list number (b)(4), that has a similar product distributed in the us with the same list number, and a 510k/pma/bla number of k173122.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Alinity I Free T4 Reagent Kit
Common Device Name
Radioimmunoassay, free thyroxine
Manufacturer (Section D)
Abbott Ireland Diagnostics Division
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
Abbott Ireland Diagnostics Division
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
+49()6122582960
MDR Report Key50275427
Report Number3005094123-2026-00399
Device Sequence Number10366755
Product Code CEC
UDI-Device Identifier00380740121822
UDI-Public(01)00380740121822(17)270112(10)83268UD00
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K173122
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07P70-30
Device Lot Number83268UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/17/2026
Supplement Report FDA Received Date08/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY I PROCESSING MODU, 03R65-01, (b)(6).
Patient SexUnknown
-
-