• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) Quadra Allure MP CRT-P; Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) Quadra Allure MP CRT-P; Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P) Back to Search Results
Model Number PM3562
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient presented remotely via merlin.Net.Upon review, the right ventricular lead exhibited noise oversensing, which resulted in pacing inhibition.However, it could not be determined whether the oversensing was due to electromagnetic interference (emi) on the pacemaker channel.No intervention was performed.There were no patient consequences.
 
Additional Manufacturer Narrative
Further information was requested but no information was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Quadra Allure MP CRT-P
Common Device Name
Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50276229
Report Number2017865-2026-17192
Device Sequence Number13365806
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public(01)05414734510097(10)A000105981(17)220831
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P030035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model NumberPM3562
Device Lot NumberA000105981
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2026
Initial Report FDA Received Date08/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1458Q/86 Quartet; 688TC/52 PKG/STER
Patient Age77 YR
Patient SexMale
-
-