| Model Number |
H74939419501510 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the mildly tortuous and severely calcified superficial femoral artery.A 5.0 x 150mm, 150cm ranger balloon catheter was advanced for dilation.During the first inflation at 6 atmospheres for 20 seconds, the balloon ruptured.The device was removed using normal method without issue and the procedure was completed with another of the same device.There were no patient complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A2 - age at time of event: 18 years or older.Investigation results:
investigation conclusion
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to patient condition.
|
| |
|