| Model Number |
D-EVOLUTFX-2329 |
| Medical Device Problem Code |
Unintended Collision (1429)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/07/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that following the successful implant of a transcatheter aortic valve, upon the attempted removal of the delivery catheter system (dcs), the nose cone caught on the frame of the valve, resulting in dislodgement.Subsequently, a second valve was implanted valve-in-valve.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: section d information references the main component of the system.Other relevant (b)(6) (s) are: product id: evfxplus-29, serial/lot #: (b)(6), ubd: 10-oct-2027, udi#: (b)(4).H6.The codes present in section h6.Correspond to multiple components/products that comprise this reported event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|