It was reported that a ventilator-dependent patient experienced acute respiratory compromise due to repeated cuff failures of bivona tight-to-shaft tracheostomy tubes.The cuffs repeatedly failed to maintain inflation, causing major air leaks, low tidal volumes, increased ventilator pressures, oxygen desaturation, agitation, and respiratory distress.Emergency tube changes and other interventions were required.Similar cuff failures occurred with multiple tubes of the same brand and size, suggesting a possible device-related issue.The problem resolved after replacing the tube with a different size, restoring effective ventilation and the patient¿s baseline condition.
|
|
This file was originally incorrectly filed under registration number mrn 9617604-2026-00191-00.The date of that submission was 10-aug-2026.One used device was returned for investigation.Device history review was performed, and no nonconformities were identified that may have contributed to the reported complaint.The visual inspection and functional testing were performed and found that the cuff failed to inflate, and a leak was observed immediately after filling.The customer complaint was confirmed.The cause of the reported issue was unknown.
|