| Model Number |
M00550350 |
| Medical Device Problem Codes |
Leak/Splash (1354); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre pulmonary dilatation balloon was being prepared to be used in the airway during an electronic bronchoscopic dilatation of trachea procedure for stenosis performed on (b)(6) 2026.During the preparation, the balloon leaked air and water.The procedure was completed with another cre pulmonary dilatation balloon.There were no patient complications reported as a result of this event.The patients condition at the conclusion of the procedure was reported to be stable.Note: the instructions for use (ifu) indicate that this balloon should not be pre-inflated prior to use in the procedure.However, the customer reported that the balloon was pre-inflated outside of the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6:
imdrf device code a0504 captures the reportable event of balloon leak in a pulmonary anatomy location.
|
| |
|