| Model Number |
CDS0801-XTW |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Unspecified Tissue Injury (4559)
|
| Date of Event |
07/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4+ functional mitral regurgitation (mr), restricted anterior leaflet for a mitraclip procedure.During the procedure, mitraclip xtw would not open.Troubleshooting was performed but was unsuccessful.Clip was caught on the atrial side of the anterior leaflet and damage was noted and clip was implanted on only anterior leaflet.Device was replaced and continued the procedure.All steps were performed as per ifu.The final mr was grade 3+.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|