Abbott Medical MitraClip G5 Delivery System; mitral valve repair devices Back to Search Results
Model Number CDS0801-XTW
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2026, a patient presented with grade 4 functional mitral regurgitation (mr), restricted posterior leaflet for a mitraclip procedure.During the procedure, clip opened 2 times after lockline removal.Troubleshooting was performed and clip was deployed.Clip opened after deployment.It was noted that clip angle was 5-10 degrees before opening and approximately 30 degrees after opening.All steps were performed as per ifu.The final mr was grade <1.There were no adverse patient effects or clinically significant delays reported.
 
Additional Manufacturer Narrative
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MitraClip G5 Delivery System
Common Device Name
mitral valve repair devices
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
Abbott Medical Northpoint Reg 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50282749
Report Number2135147-2026-05643
Device Sequence Number9753338
Product Code NKM
UDI-Device Identifier05415067050779
UDI-Public(01)05415067050779(17)271015(10)60416A1027
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCDS0801-XTW
Device Catalogue NumberCDS0801-XTW
Device Lot Number60416A1027
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2026
Initial Report FDA Received Date08/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/17/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Steerable guide catheter
Patient Age76 YR
Patient SexFemale
-
-