| Model Number |
CDS0801-XTW |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4 functional mitral regurgitation (mr), restricted posterior leaflet for a mitraclip procedure.During the procedure, clip opened 2 times after lockline removal.Troubleshooting was performed and clip was deployed.Clip opened after deployment.It was noted that clip angle was 5-10 degrees before opening and approximately 30 degrees after opening.All steps were performed as per ifu.The final mr was grade <1.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|