| Catalog Number |
00588001601 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Swelling/ Edema (4577); Joint effusion (4849)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a left total knee arthroplasty.Approximately two (2) years and two (2) months post-implantation, the patient sustained a torsional injury to the implanted joint, thus resulting in instability.Radiographic evaluation revealed displacement of the hinge mechanism indicating possible implant failure of the hinge assembly.Subsequently, the patient underwent revision surgery in which the femoral hinge post was found to be fractured.The affected components were replaced without reported complication.Due diligence is in progress for this event; to date all available information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: medical product: posterior femoral augment block precoat size f 5 mm augment with screw: catalog#: 00599003601; lot#: 64516927; distal femoral augment block precoat size f 10 mm augment with screw: catalog#: 00599003620; lot#: 64716294; qty# (b)(4); posterior femoral augment block precoat size f 10 mm augment with screw: catalog#00599003602; lot#64200187; tibial full block augment precoat size 6 10mm thickness: catalog#: 00588000610; lot#: 64973933; trabecular metal tibial cone; medium 31x31mm; catalog#: 00545001331; lot#: 65512476; trabecular metal femoral diaphyseal cone, 30mm, small, left: catalog#: 00545001730; lot#: 66362719; size 6 precoat cemented tibial component: catalog#: 00588000600; lot#: 65091872; 15mm diameter 100mm length straight stem extension combined length 145mm: catalog#: 00598801015; lot#: 65152108; 16mm diameter 100mm length straight stem extension combined length: catalog#: 00598801016; lot#: 65959358; all poly patella standard cemented size 35mm diameter 9.0mm thickness: catalog#: 00597206535; lot#: 66673129; 12mm height size f articular surface with hinge post extension: catalog#00588006012, lot#: 65049626.G2: foreign: germany
h6: proposed component code: mechanical (g04) - femur.Diligence is in progress to determine whether the device is available for further evaluation.The investigation is in progress.Upon completion of the investigation, a follow-up mdr will be submitted.
|
| |
|
Search Alerts/Recalls
|