BOSTON SCIENTIFIC NEUROMODULATION CORPORATION Vercise Genus?; stimulator, electrical, implanted, for parkinsonian tremor
|
Back to Search Results |
|
| Model Number |
DB-1232 |
| Medical Device Problem Codes |
Charging Problem (2892); Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the deep brain stimulation (dbs) patient underwent an implant procedure to implant new leads however, prior to the procedure the patient experienced difficulty charging the implantable pulse generator (ipg) out of hibernation mode and thus the ipg would not communicate with the remote control.Therefore, this ipg was explanted from the abdomen and replaced with a new ipg that was placed in the chest subcutaneously.As the lead extensions were locked into the ipg and unable to be removed, they were cut and a portion of the lead extensions remain implanted.There were no reported patient complications and the patient is stable postoperatively.
|
| |
|
Additional Manufacturer Narrative
|
Block d2b: additional pro code selection that applies to the indication of this device
additional product(s) in device system:
model number/catalog number: db-3128-95
serial number: (b)(6),
batch/lot number: n/i,
model/catalog description: dbs 2x8 extension 95cm sterile kit,
unique identifier (udi) #: (b)(4).Model number/catalog number: db-3128-95,
serial number: (b)(6),
batch/lot number: n/i,
model/catalog description: dbs 2x8 extension 95cm sterile kit,
unique identifier (udi) #: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|