Axonics Modulation Technologies, Inc. Axonics; Stimulator, Electrical, Implantable, For Incontinence
|
Back to Search Results |
|
| Model Number |
4101 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Implant Pain (4561)
|
| Date of Event |
06/15/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with this neurostimulator (ins) experienced migration of the ins and sticking out against the skin.This was causing the patient pain.Per the physician, it was possibly due to the way the patient moved when they took their prosthetic leg on and off.The physician considered planting the ins deeper in the pocket.The local care team reached out to the product management team to consult on what the maximum depth of the ins placement should be and still able to communicate with the remote control.The patient was still getting great symptom improvement and did not want to lose their device.A pocket revision surgery was scheduled at a later date.Additionally, the local care team member confirmed the ins was not coming through the skin, breaking skin, nor coming through the incision.Later, the patient had surgery for a pocket revision.No complications occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d2b:
product code: additional product code qon.Block g4:
premarket / 510(k) #: additional premarket / 510(k) # p190006.Block h11:
investigation details:
the device was not returned for analysis; therefore, a technical analysis could not be performed.Good faith effort attempts were made to try and retrieve the device; however, the product is still implanted.It was confirmed the device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.A risk review was completed and confirmed that the event of implant pain and migration was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|
Search Alerts/Recalls
|
|
|