It was reported to boston scientific corporation that an overstitch 2-0 polypropylene suture was used to treat obesity during an endoscopic sleeve gastroplasty procedure performed on (b)(6) 2026.During the procedure, the device exhibited a condition where the suture snapped when removing slack.The procedure was completed with another device of the same model.The patient is stable following the procedure, and there were no problems reported as a result of this event.
|
|
Block h6:
device code a0401 is being used to capture the reportable event of suture break.Block h11: device evaluation
the device was disposed by customer; therefore, a technical analysis could not be performed.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A similar complaint review was completed.There were no emerging trends identified that warrant further action.Boston scientific concludes that, given the event description and information provided by the customer there is not enough evidence to determine whether the suture break was due to the physician's technique during the procedure or was related to a device malfunction.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|