| Model Number |
2120 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device may have over-delivered adrucil.The patient expected a 2-day infusion; however, the device was labeled and programmed for a 24-hour infusion at 4.6 ml/hr.The pump was found empty after approximately 1 day, with 0 ml remaining.The patient was advised to contact the physician immediately due to the discrepancy in infusion duration.There was a patient involvement and there were no additional adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
|
| |
|