Model Number 72404127
Medical Device Problem Codes Migration (4003); Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this artificial urinary sphincter (aus) pump component was too high up during the healing process and the patient could not reach the pump.A surgical procedure was performed during which the device was repositioned.Upon explant the physician identified an air bubble in the pump and there was no leak in the system.The physician suspected the air bubble was potentially attributed to the previous surgery during pump component preparation.No additional information was provided.
 
Additional Manufacturer Narrative
B3: date is unknown, so estimated date was entered.Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 1100881936.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72404130.Serial number: n/a.Batch/lot number: 1100623039.Model/catalog description: belt cuff fgs 4.0 cm iz.Unique identifier (udi) #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 Artificial Urinary Sphincter
Common Device Name
Device incontinence mechanical/hydraulic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
+1(508)652-1217
MDR Report Key50288810
Report Number2124215-2026-44175
Device Sequence Number9678407
Product Code EZY
UDI-Device Identifier00878953003078
UDI-Public(01)00878953003078(17)280215(10)1100901449
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P000053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72404127
Device Catalogue Number72404127
Device Lot Number1100901449
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2026
Supplement Date Received by Manufacturer08/27/2026
Initial Report FDA Received Date08/17/2026
Supplement Report FDA Received Date09/10/2026
Date Device Manufactured02/15/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexMale
-
-