| Model Number |
72404127 |
| Medical Device Problem Codes |
Migration (4003); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this artificial urinary sphincter (aus) pump component was too high up during the healing process and the patient could not reach the pump.A surgical procedure was performed during which the device was repositioned.Upon explant the physician identified an air bubble in the pump and there was no leak in the system.The physician suspected the air bubble was potentially attributed to the previous surgery during pump component preparation.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date is unknown, so estimated date was entered.Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 1100881936.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72404130.Serial number: n/a.Batch/lot number: 1100623039.Model/catalog description: belt cuff fgs 4.0 cm iz.Unique identifier (udi) #: (b)(4).
|
| |
|