• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. VISERA CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. VISERA CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-V2
Medical Device Problem Codes Detachment of Device or Device Component (2907); Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The final investigation has been completed.The failure identified during the inspection has been confirmed.A root cause could not be identified.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
The customer returned the cysto-nephro videoscope for a separate issue.During the initial evaluation of the device it was observed, chipped light guide lens glue and rust at the opening of the channel.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA CYSTO-NEPHRO VIDEOSCOPE
Common Device Name
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50290400
Report Number3002808148-2026-28690
Device Sequence Number9742401
Product Code FAJ
UDI-Device Identifier04953170411199
UDI-Public04953170411199
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K221683
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCYF-V2
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/31/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/03/2026
Initial Report FDA Received Date08/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-