• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) UltiPace¿ Lead, 46cm MRI; Drug Eluting Permanent Right Ventricular (Rv) Or Right Atrial (Ra) Pacemaker Electrodes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) UltiPace¿ Lead, 46cm MRI; Drug Eluting Permanent Right Ventricular (Rv) Or Right Atrial (Ra) Pacemaker Electrodes Back to Search Results
Model Number LPA1231/46
Medical Device Problem Codes Failure to Capture (1081); Over-Sensing (1438); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented to hospital for an implant procedure.Post implant procedure, it was noted that the right atrial (ra) lead was dislodged resulting in no atrial capture and inappropriately sensing ventricular event.The ra lead dislodgement was confirmed with fluoroscopy.The ra lead was explanted and replaced.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UltiPace¿ Lead, 46cm MRI
Common Device Name
Drug Eluting Permanent Right Ventricular (Rv) Or Right Atrial (Ra) Pacemaker Electrodes
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50290665
Report Number2017865-2026-17275
Device Sequence Number9694774
Product Code NVN
UDI-Device Identifier05415067046345
UDI-Public(01)05415067046345(10)P000233791(17)280531
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P960013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLPA1231/46
Device Catalogue NumberLPA1231/46
Device Lot NumberP000233791
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2026
Initial Report FDA Received Date08/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Assurity MRI Pacemaker.; UltiPace Lead.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
Patient SexFemale
-
-