| Catalog Number |
SW11678 |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that no readings occurred.Data was provided for investigation.Confirmation of the complaint was confirmed via data.However, signal loss over an hour was found in connection to the allegation.The probable cause was the transmitter and app were unable to establish a connection.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Search Alerts/Recalls
|