• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. Bard, Inc. (BASD) -3006260740 PowerPort M.R.I. Implantable Port, ChronoFlex Single-Lumen, 8F; Port & Catheter, Implanted, Subcutaneous, Intravascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. Bard, Inc. (BASD) -3006260740 PowerPort M.R.I. Implantable Port, ChronoFlex Single-Lumen, 8F; Port & Catheter, Implanted, Subcutaneous, Intravascular Back to Search Results
Catalog Number 1808000
Medical Device Problem Code Expulsion (2933)
Health Effect - Clinical Codes Pain (1994); Skin Erosion (2075)
Date of Event 07/29/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported through the litigation process that, sometime after the port placement, the port allegedly malfunctioned and became exposed.Additionally, the patient experienced pain at the port site.However, the current status of the patient is unknown.
 
Additional Manufacturer Narrative
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this lot.Investigation summary: the physical device was not returned for evaluation.No medical records were provided for review.The investigation is inconclusive for the reported pain, port malfunction and exposed port issues as no objective evidence was provided for review.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PowerPort M.R.I. Implantable Port, ChronoFlex Single-Lumen, 8F
Common Device Name
Port & Catheter, Implanted, Subcutaneous, Intravascular
Manufacturer (Section D)
C.R. Bard, Inc. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
Bard Reynosa S.A. de C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key50293436
Report Number3006260740-2026-05817
Device Sequence Number9705610
Product Code LJT
UDI-Device Identifier00801741000430
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K063377
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Catalogue Number1808000
Device Lot NumberREKN0742
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2026
Initial Report FDA Received Date08/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-