• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE PRO; Catheter, Biliary, Diagnostic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE PRO; Catheter, Biliary, Diagnostic Back to Search Results
Model Number M00558710
Medical Device Problem Codes Burst Container or Vessel (1074); Inflation Problem (1310); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
Note: this report pertains to one of two cre pro wire guided dilatation balloons used in the same patient and procedure.It was reported to boston scientific corporation that two cre pro wire guided dilatation balloons were attempted to be used during a procedure performed on (b)(6) 2026.During the procedure, the balloon would not inflate due to a hole.Additionally, the same problem occurred with the other cre pro wire guided dilatation balloon.The procedure was completed using another cre pro wire guided dilatation balloon.There were no patient complications reported as a result of this event and the patient condition following procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a041001 captures the reportable event of a balloon pinhole in an unknown anatomy location.Block d4, h4: detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.Block h11: investigation results the device was not returned for analysis and was reported to be diposed; therefore, a technical analysis could not be performed.With all the available information, boston scientific concludes that the reported event of balloon pinhole could not be confirmed.The device was not returned for analysis because it was reported to be disposed.Without analysis of the actual device, there is a lack of objective evidence, or descriptive conditions of the event required to determine a probable root cause of the event.An investigation to address this problem is in progress.A review of the manufacturing documentation for this device was unable to be performed as the lot number is unknown.However, a ship history review was performed to identify the most probable lots, and a manufacturing review of the most probable lots did not identify any anomalies or deviations that could have contributed to the event.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE PRO
Common Device Name
Catheter, Biliary, Diagnostic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Cork Limited
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
+1(914)571-6918
MDR Report Key50293541
Report Number3005099803-2026-03375
Device Sequence Number10733095
Product Code FGE
UDI-Device Identifier08714729797593
UDI-Public(01)08714729797593
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K112994
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00558710
Device Catalogue NumberM00558710
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2026
Supplement Date Received by Manufacturer08/24/2026
Initial Report FDA Received Date08/18/2026
Supplement Report FDA Received Date09/17/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-