| Model Number |
8606500 |
| Medical Device Problem Codes |
Gas Output Problem (1266); Failure to Deliver (2338); Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device posted a "ventilator failure" alarm during use.There was no injury reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the device posted a "ventilator failure" alarm during use.There was no injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.Results will be provided in a separate follow-up-report.
|
| |
|
Additional Manufacturer Narrative
|
|
The ventilator failure described by the user could be reconstructed by means of the log file analysis.No indications of a device malfunction were found.Root cause for the reported symptom was an overpressure at the patient end of the circuit leading to a pressure peak.It can be assumed that a patient coughing against the ventilator caused the pressure peak.To prevent damage, the system is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.Manual ventilation and the monitoring functions remain available to the full extent.Dräger finally concludes that the device responded as specified upon the detected situation with an autonomous shutdown while changing mode to man/spont (safety mode) accompanied by an audible and visible "ventilator fail" alarm.The results of the investigation did not reveal any new risks which are not covered by the product risk management file.
|
| |
|