Datascope Corp. YAMATO PLUS-R 30cc IAB WITH ACCESSORIES AND RETRACTION KIT; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0684-00-0559-01
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2026
Type of Reportable Event Other
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: b1, b4, d10, d9, g3, g4 (pma 510k), g6, g7, h2, h6 (type of investigation, investigation findings, component codes, and investigation conclusions), h11 corrected fields: d8 (revert this field to blank, as it was inadvertently submitted initially) since the product was not returned, we were unable to evaluate the device.Potential causes of the reported issue could be an obstruction within the inner lumen or restriction (kink) within the iab catheter, guidewire and/or inner lumen that prevented passage of the guide wire.The fact that two different 0.025-inch guidewires became stuck within the same catheter, while an alternative manufacturer's catheter functioned normally under the same procedural conditions, suggests the issue was likely specific to the affected device.Based on the available information and without the return of the device we are unable to confirm the exact cause of the reported failure mode of not being able to to insert the guidewire though the iab.No decon or visual inspection performed since product was not returned and could not be evaluated.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend.Based on the rationality provided above, no escalation to the capa process is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
YAMATO PLUS-R 30cc IAB WITH ACCESSORIES AND RETRACTION KIT
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50294387
Report Number2248146-2026-0001414
Device Sequence Number15876598
Product Code DSP
UDI-Device Identifier10607567107585
UDI-Public(01)10607567107585(17)260807(10)3000327935(11)230807
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K120868
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/07/2026
Device Model Number0684-00-0559-01
Device Catalogue Number0684-00-0559-01
Device Lot Number3000327935
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2026
Supplement Date Received by Manufacturer09/16/2026
Initial Report FDA Received Date08/18/2026
Supplement Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Unknown
Patient Age62 YR
Patient SexFemale
Patient Weight88 LB
-
-