| Catalog Number |
AB35W09040080 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported during a procedure using an evercross pta balloon catheter for the treatment of a plaque lesion in the distal section of the right and left artery of the distal aorta, the balloon burst "balloon burst at, below, or unknown rated burst pressure", sp ecifically, a pin hole, which occurred on the first inflation.A 6x45 (terumo) destination sheath was used.Embolic protection was not used.The vessel was not pre or post dilated.The device did not pass through a previously deployed stent.No resistance was encountered when advancing the device.The balloon was inflated with an inflation device and 50/50 contrast was used.No damage was noted to the packaging and no issues were noted when removing the device from the hoop/tray.The instructions for use were followed without issue.It is not known at what pressure the balloon burst.The balloon did not detach during the event.The device was safely removed from the patient and the procedure was aborted.The patient was reported as alive, with no injury.No health consequences or impact were reported as a result of this event.There was no injury or intervention associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis a visual inspection of the returned device found a longitudinal tear on the balloon, the customer complaint can be confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|