Catalog Number AB35W09040080
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/03/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during a procedure using an evercross pta balloon catheter for the treatment of a plaque lesion in the distal section of the right and left artery of the distal aorta, the balloon burst "balloon burst at, below, or unknown rated burst pressure", sp ecifically, a pin hole, which occurred on the first inflation.A 6x45 (terumo) destination sheath was used.Embolic protection was not used.The vessel was not pre or post dilated.The device did not pass through a previously deployed stent.No resistance was encountered when advancing the device.The balloon was inflated with an inflation device and 50/50 contrast was used.No damage was noted to the packaging and no issues were noted when removing the device from the hoop/tray.The instructions for use were followed without issue.It is not known at what pressure the balloon burst.The balloon did not detach during the event.The device was safely removed from the patient and the procedure was aborted.The patient was reported as alive, with no injury.No health consequences or impact were reported as a result of this event.There was no injury or intervention associated with this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product analysis a visual inspection of the returned device found a longitudinal tear on the balloon, the customer complaint can be confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERCROSS 035
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
Covidien
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
Covidien
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
+353(091)708096
MDR Report Key50297095
Report Number2183870-2026-00393
Device Sequence Number18596391
Product Code LIT
UDI-Device Identifier00763000767037
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K082579
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberAB35W09040080
Device Lot NumberC130497
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2026
Initial Report FDA Received Date08/18/2026
Date Device Manufactured06/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age39 YR
Patient SexFemale
Patient Weight128 LB
-
-