| Model Number |
M001491561 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
Unspecified Tissue Injury (4559)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Ecn event description: ablations performed in left pulmonary veins, no issues noted.Physician having issues attempting to wire rpvs.What troubleshooting steps took place? what troubleshooting steps, if any, resolved the issue?: deployed 'flower' conformation when farawave catheter partially in sheath.Non-recoverable.What is the next course of action?: new farawave catheter opened.Patient present at time of event?: yes.
|
| |
|
Event or Problem Description
|
|
Event description: ecn event description: ablations performed in left pulmonary veins, no issues noted.Physician having issues attempting to wire rpvs.What troubleshooting steps took place? what troubleshooting steps, if any, resolved the issue?: deployed 'flower' conformation when farawave catheter partially in sheath.Non-recoverable.What is the next course of action?: new farawave catheter opened.Patient present at time of event?: yes.Event date: (b)(6) 2026.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history records for the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.Awaiting device return.
|
| |
|
Additional Manufacturer Narrative
|
|
No device was returned for analysis.A review of the device history records for the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.The complaint is non-verifiable as the product was not returned for evaluation.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided by the supporting documentation, "emerging use" appears to have impacted on the event as reported.Without the return of the actual device in question for evaluation, lake region medical is unable to determine the exact cause for this incident.Lake region medical has no open preventative action specific to manufacturing this product differently.As indicated in the ifu (instructions for use) precautions section:
· prior to use, all equipment to be used for the procedure should be carefully examined to verify proper function and integrity.
|
| |
|