(b)(4).Date sent 8/18/2026.D4 batch # 841d78.Investigation summary.The product was returned for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample revealed that the cdh25a device arrived with no apparent damaged.The breakaway washer was present and uncut, all staples were present and protruding inside the pockets.Further analysis shows marks on the safety.No functional test was performed to preserve evidence.It appears that an attempt was made to fire the device with the safety engaged.It should be notice that to fire the device, draw the red safety back toward the adjusting knob until it seats into the body of the device.If the safety cannot be released, the device is not in the safe firing range.Please reference the instructions for use for additional information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 841d78, and no non-conformances were identified.
|