This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, h2, h3, h4, h6 and h11.The device was returned to the regional service center for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: noise image.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: regarding the customer¿s allegation, since the scope communication error b30 due to the printed circuit board had become loose., the locking mechanism has become loose, causing noise in the image, the cause was traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|