Model Number 101045439
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Code Inflammation (1932)
Date of Event 05/16/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Customer service sent a replacement pump to the customer and asked for the return of her old pump for evaluation.In follow up with a complaint handler on (b)(6) 2026, the customer stated that she was sent a replacement pump.The customer stated she had developed mastitis on (b)(6) 2026 and after a week her mastitis had resolved.The customer was sent a return label for her original magic inbra double pump, the return label has not been used.Based on the results of our internal investigation (capa-2025-0027) including published literature, it cannot definitively be concluded that the pump caused or has contributed to the customer's reported mastitis.According to published clinical literature, mastitis incidence in lactating women, with or without a breast pump, can be as high as 33%.In fact, clinical guidelines recommend the use of a breast pump to facilitate the removal of breast milk during periods of mastitis.Additionally, complaint trends across all medela pumps show mastitis incidence consistently below 0.1%, far lower than the 3-20% occurrence in the general population of lactating women "[2]".Customer service sent a replacement pump to the customer and requested the old pump back for evaluation and testing.Multiple attempts to follow up with the customer by a complaint handler have gone unanswered as of the date of this report.Based on the results of our internal investigation (capa-2025-0027) including published literature, it cannot definitively be concluded that the pump caused or has contributed to the customer's reported mastitis.According to published clinical literature, mastitis incidence in lactating women, with or without a breast pump, can be as high as 33%.In fact, clinical guidelines recommend the use of a breast pump to facilitate the removal of breast milk during periods of mastitis.Additionally, complaint trends across all medela pumps show mastitis incidence consistently below 0.1%, far lower than the 3-20% occurrence in the general population of lactating women "[2]".
 
Event or Problem Description
On (b)(6) 2026, the customer contacted medela llc that while using the magic inbra she was diagnosed with mastitis due to poor or insufficient suction.Additionally, she alleged that she was prescribed antibiotics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Magic InBra Double Pump EN FR ES
Common Device Name
Pump, Breast, Powered
Manufacturer (Section D)
Medela LLC
1101 corporate dr
mchenry IL 60050
Manufacturer (Section G)
Medela AG
lattichstrasse 4b
baar zug
SZ  
Manufacturer Contact
alea raber
1101 corporate drive
mchenry, IL 60050
+1(800)435-8316
MDR Report Key50300672
Report Number1419937-2026-00021
Device Sequence Number21513789
Product Code HGX
UDI-Device Identifier07612367087599
UDI-Public07612367087599
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K251377
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 08/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number101045439
Device Catalogue Number101045439
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/17/2026
Initial Report FDA Received Date08/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-