| Model Number |
1-2-510.003 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that the door to their advantage plus endoscope reprocessing system opened mid cycle and sprayed water.No report of injury.
|
| |
|
Event or Problem Description
|
|
The user facility reported that the door to their advantage plus endoscope reprocessing system opened mid cycle and sprayed water.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the reported event is in process.A follow-up report will be submitted once additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite following the reported event to inspect the unit.The reported event could not be duplicated.To resolve the issue, the technician replaced the scu control board and the damaged temperature probe.The technician tested the unit, confirmed it to be operating according to specification, and returned it to service.The device history record was reviewed and confirmed the device was manufactured to specifications; no abnormalities were noted.A 3-year complaint review confirmed this to be an isolated event.No additional issues have been reported.
|
| |
|