Esi-electrosurgical instrumest salvati proctoscope 15-4010: multiple product (2)
arrived new, was reprocessed once each using manual cleaning with fda registered
cleaning agent prepared per ifu and aami water standards; sent through fda
registered washer/disinfector using a fda registered cleaning agent prepared per ifu.All device reprocessing instructions followed as required per the manufacturer ifu.Product was then packaged and sterilized with a steris fda registered autoclave at
the pre-vacuum steam sterilization parameters per the device ifu at 270*f, 4-minute
exposure.Once completed product was opened to be used for a patient procedure and
was found to be defective.The salvati proctoscope consists of 2 pieces, an inner
plastic piece and an outer metal piece.After reprocessing 1-time, the inner plastic
piece could not be assembled into the outer metal piece and unable to be used.The
surgeon had to used rectal retractor's instead creating a less than ideal surgical
situation.No harm was done and procedure was completed.A second salvati was
purchased new and the exact same issue occurred.Both salvati proctoscopes could
not be used due to the inner plastic piece not fitting into the outer metal sleeve.Devices
unable to be used.Inner plastic piece fit into the outer metal piece prior to
reprocessing, all device reprocessing steps followed per ifu, prolystica ph neutral
enzymatic detergent, the getinge washer/disinfector, and the steris amsco steam
sterilizer were used per each devices ifu.Issue was reported to device manufacturer.Reference report mw6001216.Annex a: 2588, 1670, 1091.Annex e: 4582.Annex f: 4642.Annex g: 4723.This report captures salvati operating proctoscope stainless #1.
|