| Catalog Number |
6212001 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
The customer reported that following a plasmapheresis procedure on (b)(6) 2026 using a rika device, the donor was found deceased on (b)(6) 2026.The collection set is not available for return because it was discarded by the customer.This report is being filed due to a reported patient death.At this time, the available information does not indicate a detectable malfunction of the terumo bct device, and no device malfunction has been alleged.
|
| |
|
Event or Problem Description
|
|
The customer reported that following a plasmapheresis procedure on (b)(6) 2026 using a rika device, the donor was found deceased on (b)(6) 2026.The collection set is not available for return because it was discarded by the customer.This report is being filed due to a reported patient death.At this time, the available information does not indicate a detectable malfunction of the terumo bct device, and no device malfunction has been alleged.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: a service technician checked out the device at the customer site and concluded that the device performed as expected with passing results. investigation is in process, a follow-up report will be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information in h.6 and h.11.Investigation: a service technician checked out the device at the customer site and concluded that the device performed as expected with passing results. the run data file (rdf) was analyzed for this event.The log file shows that over the 37-minute procedure, there was one 4602 access pressure too low alarm in the 6th draw cycle, which the operator was able to continue.The ac fluid detector, donor line fluid detector, line sensor, aps, return cps, and draw cps signals were analyzed, and nothing anomalous was found. a review of the device history record (dhr) for this lot showed no irregularities during manufacturing that were relevant to this issue.The history of the separation set lot 2604243161 and plasma bottle 2603041002 were reviewed and there have not been any recalls for these lots.Investigation is in process, a follow-up report will be provided.
|
| |
|
Search Alerts/Recalls
|