| Model Number |
PDVXL6 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/24/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient was implanted with a prodisc c vivo device on (b)(6)2025.It is unknown if the patient experienced any symptoms, however, imaging showed that the implant had migrated.The implant was removed on (b)(6)2026 and the patient was fused.
|
| |
|
Additional Manufacturer Narrative
|
|
An mdr is indicated for this complaint.A patient was implanted with a prodisc c vivo device on (b)(6)2025.It is unknown if the patient experienced any symptoms, however, imaging showed that the implant had migrated.The implant was removed on (b)(6)2026 and the patient was fused.A dhr review found no anomalies associated with the complaint.Complaint trending found that the rate of complaints is within the level outline in the risk documentation.A review of the risk documentation found that the hazards associated with the complaint are identified and mitigated to a level where the clinical benefits outweigh the harms.The implant was not returned following removal surgery therefore, no device evaluation could be completed.Imaging was provided that showed the device had migrated.There were no anomalies identified during the complaint investigation.The removal was completed due to implant migration.The submission is 1 of 1 devices involved in this event.
|
| |
|