• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) Entrant HF; Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) Entrant HF; Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D) Back to Search Results
Model Number CDHFA300Q
Medical Device Problem Codes Over-Sensing (1438); Pacing Asynchronously (1441)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that post-paced t-wave oversensing (pptwos) was noted on the device.Abbott technical support was contacted and reprogramming was recommended.No intervention has been performed at this time.The patient was in stable condition and will continue to be monitored.
 
Event or Problem Description
Additional information was received indicating the post-paced t-wave oversensing (pptwos) noted was due to fusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Entrant HF
Common Device Name
Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50306094
Report Number2017865-2026-17332
Device Sequence Number10124856
Product Code NIK
UDI-Device Identifier05415067032850
UDI-Public(01)05415067032850(10)P000153231(17)240831
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model NumberCDHFA300Q
Device Catalogue NumberCDHFA300Q
Device Lot NumberP000153231
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/18/2026
Supplement Report FDA Received Date08/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-