• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Inc. STRATAFIX; Suture, surgical, absorbable, polydioxanone

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Inc. STRATAFIX; Suture, surgical, absorbable, polydioxanone Back to Search Results
Catalog Number SXPP1A405
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/17/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent a single port laparoscopic abdominal myomectomy, pelvic adhesiolysis, and diagnostic curettage under general anesthesia on (b)(6) 2026 and a barbed suture was used.Clinical diagnosis: 1.(multiple) uterine leiomyoma (type v) 2.Female pelvic adhesions.3.Abnormal uterine bleeding 4, (left adnexal area cystic mass) ovarian cyst 5, moderate anemia during the procedure, the surgeon used suture to stop bleeding on the uterine wound.Operate was normal according to surgical specifications.During use, the needle that came with the suture suddenly broke and fell into the abdominal cavity.Due to the needle remaining in the abdominal cavity, additional surgical exploration procedures were required, which prolonged the surgical duration and increased the risk of abdominal exposure, tissue damage, and infection.After the doctor discovered the broken needle, immediate measures were taken: abdominal exploration was performed immediately during the operation.Due to severe pelvic adhesion in the patient, it was difficult to find it under laparoscopy, so the surgery was converted to open surgery.After careful search, the broken needle was successfully removed.Subsequently, the patient received infection prevention and symptomatic supportive treatment after surgery.The patient's vital signs remained stable without any complications such as residual foreign bodies, bleeding, or infection in the abdominal cavity.Currently, the patient's condition has improved and is recovering.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.What is the surgeon¿s experience with this stratafix suture? please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure.On what tissue was the suture used? what was the tissue condition (normal, thin, calcified, fragile, diseased)? what instruments were used to grasp the needle? where was the needle grasped during use? were x-rays performed to localize the needle fragment? what was the size of the broken needle fragment? where did the needle break (tip, middle, swage area)? did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? other relevant patient history/concomitant medications? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status?.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX
Common Device Name
Suture, surgical, absorbable, polydioxanone
Manufacturer (Section D)
Ethicon Inc.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
Ethicon Inc.-Juarez
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
+1(303)552-6892
MDR Report Key50306651
Report Number2210968-2026-08353
Device Sequence Number13434006
Product Code NEW
UDI-Device Identifier10705031227682
UDI-Public(01)10705031227682(17)270228(10)107A23
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
K182873
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSXPP1A405
Device Lot Number107A23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2026
Supplement Date Received by Manufacturer08/19/2026
09/21/2026
Initial Report FDA Received Date08/18/2026
Supplement Report FDA Received Date09/15/2026
10/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/25/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-