Event or Problem Description
|
|
A complaint was received n which the customer reported an intraoperative incident involving item 103.619 (multifluted zygoma-s bur, lot ykwx5).According to the practitioner's report, the multifluted bur fractured, resulting in the bur becoming lodged within the patient's left maxillary sinus.However, the product safety management and product research & development departments conducted the relevant investigations and concluded that the item's drilling guide pin had unscrewed and detached.
|
| |
|
Additional Manufacturer Narrative
|
|
Post market surveillance data confirms that the benefit-risk ration maintains acceptable and will keep investigating occurrences related to this product.An internal capa will be started to feedback risk management as necessary, prepare pms process to track new occurrences and check for improvements opportunities in production process or design related.Non-fatal serious injury or device malfunction stored due to covid 19 pandemic in accordance with fda guidance "postmarketing adverse event reporting for medical products and dietary supplements during a pandemic" published may 2020.
|
| |
|