• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HD AUTOCLAVABLE CAMERA HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HD AUTOCLAVABLE CAMERA HEAD Back to Search Results
Model Number CH-S190-XZ-E
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported problem with image during service maintenance involving an olympus camera head.Additionally, during the device analysis, intermittent image and failed leak test were observed.There was no patient injury reported.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: intermittent image and failed leak test.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction problem with image was traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD AUTOCLAVABLE CAMERA HEAD
Common Device Name
AUTOCLAVABLE CAMERA HEAD
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50307500
Report Number3002808148-2026-28855
Device Sequence Number10630145
Product Code FET
UDI-Device Identifier04953170310799
UDI-Public04953170310799
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K102059
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCH-S190-XZ-E
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/30/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 07/27/2026
Initial Report FDA Received Date08/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-