• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION CoverEdge? 32; Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION CORPORATION CoverEdge? 32; Stimulator, Spinal-Cord, Totally Implanted For Pain Relief Back to Search Results
Model Number SC-8336-50
Medical Device Problem Codes Unexpected Therapeutic Results (1631); Migration (4003)
Health Effect - Clinical Code Pain (1994)
Date of Event 06/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the lead migrated which was confirmed with imaging and the patient was also noted to be experiencing inadequate pain relief.The patient underwent paddle repositioning procedure.Patient had great coverage postoperatively.
 
Additional Manufacturer Narrative
Investigation result: the device was not returned to boston scientific for analysis.Device history record: a review of the device history record (dhr) was conducted.The records indicate that the device met all applicable acceptance criteria and release requirements prior to distribution.No manufacturing-related nonconformances or deviations were identified during the review that could be associated with the reported event.Labeling review: a labeling review did not reveal any anomalies as it states: lead migration, resulting in undesirable changes in stimulation and subsequent reduction in pain relief.System failure, which can occur at any time due to random failure(s) of the components or the battery.These events, which may include device failure, lead breakage, hardware malfunctions, loose connections, electrical shorts or open circuits and lead insulation breaches, can result in ineffective pain control and undesirable stimulation may occur over time due to cellular changes in tissue around the electrodes, changes in electrode position, loose electrical connections and/or lead failure.Investigation conclusion: based on a thorough review of the reported complaint, an investigation conclusion code of cause not established was assigned to the investigation for all components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CoverEdge? 32
Common Device Name
Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
Boston Scientific Corporation
road 698, lot no. 12
dorado PR 00646 -260
PR   00646-2602
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
+1(763)292-0920
MDR Report Key50308185
Report Number3006630150-2026-06021
Device Sequence Number18800376
Product Code LGW
UDI-Device Identifier08714729832669
UDI-Public(01)08714729832669(17)271014(10)7093227
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberSC-8336-50
Device Catalogue NumberSC-8336-50
Device Lot Number7093227
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2026
Initial Report FDA Received Date08/18/2026
Date Device Manufactured10/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
-
-